Abstract
The purpose of this experimental study was to test the reliability of a single malleable
titanium miniplate using Champy's method of fixing fractures of the mandibular angle.
Eighteen sheep hemimandibles were used to evaluate 2 plating techniques. The groups
were tested with either a single non-compression titanium miniplate or a single malleable
titanium miniplate. A cantilever bending biomechanical test model was used for the
samples. Each group was tested with vertical forces using a servohydraulic testing
unit. The displacement values in each group at each 10 N stage up to 90 N were compared using 2-way analysis of variance (ANOVA). The displacement values
for the 2 groups differed significantly (p < 0.01). The variance analyses showed that the biomechanical behaviour of a single non-compression
miniplate was better than that of a single malleable miniplate. The non-compression
miniplate fixed by screws had greater resistance to occlusal loads than the malleable
plate fixed by screws, and the malleable plate alone was not sufficient to withstand
the early postoperative bite force.
Keywords
To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to British Journal of Oral and Maxillofacial SurgeryAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- A retrospective analysis of 327 mandibular fractures.J Oral Maxillofac Surg. 1983; 41: 305-309
- Ten years of mandibular fractures: an analysis of 2,137 cases.Oral Surg Oral Med Oral Pathol. 1985; 59: 120-129
- An epidemiologic survey of facial fractures and concomitant injuries.J Oral Maxillofac Surg. 1990; 48: 926-932
- Treatment of mandibular angle fractures using one noncompression miniplate.J Oral Maxillofac Surg. 1996; 54: 864-871
- Treatment of mandibular fractures by means of osteosynthesis without intermaxillary immobilization according to F.X. Michelet's technic (in German).Zahn Mund Kieferheilkd Zentralbl. 1975; 63: 339-341
- Mandibular osteosynthesis according to the Michelet technic. II. Presentation of new material. Results (in French).Rev Stomatol Chir Maxillofac. 1976; 77: 577-582
- Mandibular osteosynthesis by miniature screwed plates via a buccal approach.J Maxillofac Surg. 1978; 6: 14-21
- Fracture line stability as a function of the internal fixation system: an in vitro comparison using a mandibular angle fracture model (discussion).J Oral Maxillofac Surg. 1995; 53: 801
- Treatment of mandibular angle fractures with a malleable noncompression miniplate.J Oral Maxillofac Surg. 1999; 57: 288-293
- Bite forces in patients treated for mandibular angle fractures: implications for fixation recommendations.J Oral Maxillofac Surg. 1994; 52: 734-736
- Bite forces in patients after treatment of mandibular angle fractures with miniplate osteosynthesis according to Champy.Int J Oral Maxillofac Surg. 2002; 31: 345-348
- Measure of bite force and occlusal contact area before and after bilateral sagittal split ramus osteotomy of the mandible using a new pressure-sensitive device: a preliminary report.J Oral Maxillofac Surg. 2000; 58: 370-374
- Prospective study of mandibular fractures.J Oral Surg. 1981; 39: 275-281
- Infection after rigid internal fixation of mandibular fractures: a clinical and radiologic study.J Oral Maxillofac Surg. 1991; 49: 585-593
- Treatment methods for fractures of the mandibular angle.Int J Oral Maxillofac Surg. 1999; 28: 243-252
- Monocortical miniplate fixation of mandibular fractures.Arch Otolaryngol Head Neck Surg. 1991; 117: 149-154
- The development of modern osteosynthesis (in German).Mund Kiefer Gesichtschir. 2000; 4: s84-s90
- Small plate osteosynthesis of mandibular fractures.Br J Oral Maxillofac Surg. 1985; 23: 77-91
- The transoral treatment of mandibular fractures using noncompression miniplates: a prospective study.Ann Plast Surg. 1988; 20: 409-413
- Monocortical noncompression miniplate osteosynthesis of mandibular angle fractures.Aust NZ J Surg. 1990; 60: 805-809
- Transoral 2.0-mm miniplate fixation of mandibular fractures plus 2 weeks’ maxillomandibular fixation: a prospective study.J Oral Maxillofac Surg. 2002; 60: 167-170
- Plating techniques and plate orientation in repair of mandibular angle fractures: an in vitro study.Laryngoscope. 1998; 108: 1218-1224
Article info
Publication history
Published online: November 28, 2011
Accepted:
October 27,
2011
Identification
Copyright